Company record
Heraeus Medical GmbH (Dental Division)
0 FDA warning letters since 2023 and 10 recalls in the last 12 months.
Get an email the day FDA posts anything new about Heraeus Medical GmbH (Dental Division).
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Oct 7 | Class II | Heraeus Medical GmbH (Dental Division) Device |
PALACOS MV+G pro 40; PALACOS MV+G pro 80. Model/Catalog Number: 5081290; 5081291. Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with … Potential for monomer liquid leakage. |
| Oct 7 | Class II | Heraeus Medical GmbH (Dental Division) Device |
PALACOS MV pro 40; PALACOS MV pro 80. Model/Catalog Number: 5150633; 5150635. Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or … Potential for monomer liquid leakage. |
| Oct 7 | Class II | Heraeus Medical GmbH (Dental Division) Device |
PALACOS R+G pro 40; PALACOS R+G pro 80. Model/Catalog Number: 5081288; 5081289. Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with o… Potential for monomer liquid leakage. |
| Oct 7 | Class II | Heraeus Medical GmbH (Dental Division) Device |
PALACOS R pro 40; PALACOS R pro 80. Model/Catalog Numbers: 5081286; 5081287. Radiopaque, poly (methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or w… Potential for monomer liquid leakage. |
| May 6 | Class III | Heraeus Medical GmbH (Dental Division) Device |
Heraeus, PALAMIX uno. Material Number: 66057893. Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months). |
| May 6 | Class III | Heraeus Medical GmbH (Dental Division) Device |
Heraeus, PALAMIX duo. Material Number: 66057897. Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months). |
| Jan 21 | Class II | Heraeus Medical GmbH (Dental Division) Device |
Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cemen… increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed. |
| Jan 21 | Class II | Heraeus Medical GmbH (Dental Division) Device |
Brand Name: PALACOS Product Name: PALACOS MV pro 40; PALACOS MV pro 80 Model/Catalog Number: 5150633; 5150635 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement,… increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed. |
| Jan 21 | Class II | Heraeus Medical GmbH (Dental Division) Device |
Brand Name: PALACOS Product Name: PALACOS R pro 40; PALACOS R pro 80 Model/Catalog Number: 5081286; 5081287 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, p… increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed. |
| Jan 21 | Class II | Heraeus Medical GmbH (Dental Division) Device |
Brand Name: PALACOS Product Name: PALACOS MV+G pro 40; PALACOS MV+G pro 80 Model/Catalog Number: 5081290; 5081291 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cem… increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed. |